High Pressure cylinders should be used in facilities that have low or intermittent requirements for medical gas. These systems are not affected by the warming of liquid cylinders and therefore do not require economizer circuits and liquid cylinder maintenance.

High Pressure cylinders should be used in facilities that have low or intermittent requirements for medical gas. These systems are not affected by the warming of liquid cylinders and therefore do not require economizer circuits and liquid cylinder maintenance.||The system shown is comprised of a 3, 5, 7 or 9 cylinder manifold on each side of a control section with automatic switches from the primary supply to the secondary supply when the pressure in the primary cylinders drop below a preset pressure.||A pressure switch is provided to send a signal to an alarm panel that indicates when the reserve cylinder bank is in use. Both cylinder banks feed the house line through a dual regulator manifold as described in the liquid cylinder system.

Markets
Applications
Connections Inlet_CGA Outlet_¼” FNPT
Flow Capacity .0-4,000 SCFH
Inlet Pressure 0-3,000 PSIG
Outlet Pressure 0-55 PSIG
Material Brass
Options Number of cylinders per header (3, 5, 7, 9) Typically, primary and secondary headers have the same number of headers
Product Type Medical Reserve Manifolds